ELPRO Libero Cx Instrucciones de operaciones Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 104
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
42 - EN
elproVIEWER
SI3001Ea
APPLICATIONS
4
4.2 Assessment
During assessment, the existing LIBERO PDF report is created again but with a different
monitoring period of different alarm thresholds. To prepare for assessment, a new project
is created containing the LIBERO PDF report requiring assessment.
2 Start of the elproVIEWER / Project
The following sections: 4.2.1.1 Scenario - Time to 4.2.2.2 Settings - threshold value each
describe an example of assessment with a different monitoring period and different alarm
thresholds.
4.2.1 Assessment - time
4.2.1.1 Scenario - Time
In the following example, the data logger had been stopped too late, i.e. at a point at which
it was no longer at refrigerated conditions (2-8°C). The alarm-triggering time range is
deleted in this assessment.
The blue section of the
line triggered the
alarm
Original PDF report at the freight recipient with alarm
Vista de pagina 41