liberoCONFIG
LI6003E
CONFIGURATION OF A LIBERO CX
3
3.4.3.3 MKT Alarm
MKT Mean Kinetic
Temperature
MKT alarm and duration alarm
MKT Activation
Energy
Possibility to enter a product specific value for the MKT activation energy between 42....
125 kJ/mol. (standard value: 83 kJ/mol.)
MKT Alarm Enable MKT alarm
A checked box activates the alarm mode.
Lower MKT Limit
Data entry field for the threshold value.
Upper MKT Limit
Data entry field for the threshold value.
Only the section "MKT Alarm" is shown on this screen. For the maximum duration
see 3.4.3.4 Duration Alarm
As a reference value for the lower MKT threshold we recommend 0 °C. This entry
only serves to complete the general specifications for threshold values (min. -
max. value).